Click here to Print this Window
Opiates Technical Information
Cross-Reactivity of IDS One-Step ELISA for Opiates
Compound
Concentration resulting in 50% inhibition (ng/ml)
Dihydrocodeine
0.175
Heroin
0.3
Codeine
0.35
Morphine
0.425
6-Monoacetylmorphine
0.5
Morphine-6-beta-glucuronide
5.5
Morphine-3-beta-glucuronide
5.5
Hydrocodone
8.5
Hydromorphone
12.5
Oxycodone
700
Oxymorphone
1000
There is a possibility that other substances and/or factors not listed above may interfere with the test and cause false results. Results may be influenced by matrix and assay conditions.

C:XL/CR-OP-Omnoadd
Rev: n/a
SS:  n/a
Cre: 52405